Re: Dead time

From: Joseph Trotter (trotter@scripps.edu)
Date: Wed Apr 28 1999 - 20:35:23 EST


Dear Sasha,

	I forgot an important point - we have 4 PMT detectors on laser 1, and
have measured PE-Cy7 off the 1st laser as well as APC-Cy7 off the
2nd laser. We use the same 780/60 BP filter for both tandem conjugates
involving Cy7 and the 710 LP dichroic as well. I have not tested a 710 SP
dichroic for Cy7 reagents. You ought to use a BP in addition to your
mirror to minimize leakage from other areas of the spectrum. We have left the
standard 610 SP in place to reflect signal to the 710LP dichroic.

		Sorry for the omission,

		Joe



This archive was generated by hypermail 2b29 : Wed Apr 03 2002 - 11:53:25 EST