Re: red sensitive PMT, small things

Mario Roederer (ROEDERER@Darwin.Stanford.EDU)
Mon, 19 Jun 1995 22:14:17 -0700 (PDT)

We have used a red-sensitive PMT, but not for Cy5 or PAC or TR... but
for a dye that's further out (at 780 nm). It improved signal to
background about 5-fold over the standard PMTs. (Immunofluorescence staining).

I have done FACS on very small things... but less than 0.2 microns!
Wow. In any case, we found that using log side scatter was a
considerably better trigger parameter than FS. In Bob Murphy's lab, we
used to centrifuge our sheath fluid at 20,000 x g to remove
particulates. This reduced the "background" event rate considerably.
Here at Stanford, I did FACS on myxococcus xanthus (an enormous 1 x 6
microns), using log side scatter--we did not have to treat the sheath
fluid specially. In both cases, the forward scatter signal was
basically meaningless. In both cases, we could easily detect
fluorescence (either fluorescein or rhodamine).

mr


Home Page Table of Contents Sponsors Web Sites
CD ROM Vol 2 was produced by staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone:(317) 494-0757; FAX (317) 494-0517; Web http://www.cyto.purdue.edu EMAIL robinson@flowcyt.cyto.purdue.edu