The examples in the paper you cited indicate CD38 intensity could be used to
identify plasma cells. Of course, it is essential to perform cytoplasmic
kappa vs. lambda light chain immunoglobulin to prove clonality in abnormal
plasma cell populations. Less than 10% of the cases of monoclonal plasma
cells match the CD38 expression on lymphoblasts (in cases of plasma cell
leukemia), and in these instances, other criteria such as CD45 and light
scatter could be used to distinguish the cells. As you know, the definition
of complex populations by a single parameter is dangerous at best.

Denise A. Wells, M.D.


Home Page Table of Contents Sponsors Web Sites
CD ROM Vol 2 was produced by staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone:(317) 494-0757; FAX (317) 494-0517; Web http://www.cyto.purdue.edu EMAIL robinson@flowcyt.cyto.purdue.edu