Eric Martz (emartz@microbio.umass.edu)
Tue, 02 Aug 1994 18:40:57 -0500 (EST)

Can you guys at Molecular Probes (or anyone else) tell me:

1. What is the fluorescence intensity ratio for emission of
calcium-saturated fluo-3 vs. fluorescein?

2. Is fluo-3 emission pH sensitive as for fluorescein?

Knowing #1 would allow me to estimate, from the molecules of equivalent
soluble fluorescein (MESF)/cell, the absolute concentration of fluo-3 in
the cell (assuming uniform distribution). In our case, we're exciting with
a 488 nm laser, and detecting in a 530 +/- 30 nm filter. But I just want a
rough idea. Assuming the emission ratio is 1, loading lymphocytes with
fluo-3 at 0.04 micromolar for 10 min, I calculate that the intracellular
concentration is about 0.2 micromolar. This gives us a signal to noise
ratio of 5 for basal cell fluorescence (no stimulus nor ionophore) where
noise is autofluorescence, i.e. quite workable. MESF is obtained by
reference to commercially available standard beads (Flow Cytometry
Standards Corp.). Fluo-3 signal goes up about 10-fold with ionophore.

/*- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Eric Martz, Professor of Immunology emartz@microbio.umass.edu
Dept Microbiology Voice: 413-545-2325 FAX: 413-545-1578
Morrill IVN 203, Box 35720, Univ Massachusetts, Amherst MA 01003-5720
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -*/


Home Page Table of Contents Sponsors Web Sites
CD ROM Vol 2 was produced by staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone:(317) 494-0757; FAX (317) 494-0517; Web http://www.cyto.purdue.edu EMAIL robinson@flowcyt.cyto.purdue.edu