Yes, they are used OK here. You don't detail the problem, so this may or may not be relevant: Tandem dyes such as PE-Cy5 are more fragile than the simpler ones, will give highly variable results if the PE and Cy5 components get partially separated. This can be caused by exposure to light &/or heat, and shows up as variations in compensation and intensity. Cheers, Joseph. At 02:35 7/11/2002, Julie Nelson wrote: >Hi ya'll, >Has anyone successfully compensated PE-Cy5 (FL3) and APC (FL4) on a >FACSCalibur? I have tried both during acquistion and post-aquisition >and have not had much luck. I am currently not recommending it, but I >am curious as to why it doesn't seem to work. >Thanks, >Julie -- Joseph Webster, Flow Cytometry Facility Centenary Institute, Sydney, AUSTRALIA. Phone +61-2-9565-6110