"Witherspoon, Sam" wrote: > > I'm trying to collect DAPI and APC signals on the second (and third) > laser lines of a FACStar Plus > The DAPI signal is very bright... should I reflect it and filter the > APC fluorescence or should I filter the DAPI and reflect the APC? Hi Sam, I recently tried to work this combo out on our Elite. Setup was 488 beam displaced from the other two beams [uv and 633] which were set colinear. An unexpected result was that DAPI fluorescence showed up in the APC signal. I tried different bandpass and longpass filters for the APC and ran samples with variuos lasers blocked to verify the souurce of the signal. My conclusion was that DA{PI simply fluoresces way out into the red, at least under the experiemental conditions I used. Regards, Tom ***************************************************************************** Thomas W. Mc Closkey, Ph. D., Director of Flow Cytometry North Shore University Hospital, New York University School of Medicine Boas Marks Biomedical Research Center, 350 Community Drive Manhasset, Long Island, New York 11030 ph: 516-562-4844 [office], 516-562-1135/4641 [lab] fax: 516-562-2866 *****************************************************************************