Re: sheath fluid for Calibur

From: Alice L. Givan (Alice.L.Givan@Dartmouth.EDU)
Date: Wed Mar 29 2000 - 17:09:38 EST


Hello Flowers,
We have found that  using distilled/de-ionized water in the sheath is noisier than
buffer on our cytometers.  After puzzling over the question of contaminants in the
water and after some discussion with a few gurus,  we have concluded that the noise
is not from contaminated water but may result from light bouncing off the refractive
index change at the water (sheath): buffer (sample core) interface.

Any thoughts on this?

Alice