Re: Microparticle Analysis Using FACScan/Sort

From: Gerhard Nebe-von-Caron (Gerhard.Nebe-von-Caron@Unilever.com)
Date: Fri Apr 09 1999 - 11:07:25 EST


          Most current commercial cytometers tail off at 500nm scatter 
          detection. trigger should be side scatter, preferential peak 
          signal.
          
          Initial strategy is to use 500nm yellow/green beads from 
          polysciences. Dilute particles to reach approx. 10^6/ml (see 
          formula in their particle catalog). Look at green 
          fluorescence versus sisdescatter. Threshold on green 
          fluorescence with your normal lymphocyte settings.Once you 
          see the fluorescent signals lower threshold down to noise 
          level. Adjust scatter to have beads in the middle of the 
          side scatter scale. Change to side scatter triggering with a 
          setting at mid scale and reduce slowly. You should 
          eventually see non-fluorescent noise signals in the low 
          scatter area. Run filtered sample liquid as a control.
          
          On the XL we routinely analyse 320 nm particles triggered on 
          peak side scatter, but it requires a clean sheath system. 
          
          Good luck
          
          Gerhard 


______________________________ Reply Separator _________________________________
Subject: Microparticle Analysis Using FACScan/Sort
Author:  laurie_stojanovic@baxter.com at INTERNET
Date:    02/04/1999 20:15


Has anyone out there used a Becton Dickinson FACScan or FACSort flow
cytometer to analyze platelet- or WBC-derived microparticles (0.2-1.0 um in
size)?  I am looking for assistance in the following arenas:

     1.  Distinguishing background (noise) events from "real" microparticle
events; looking for ways to increase the limit of            sensitivity of
these instruments (Becton Dickinson Suggests limit is 0.5 to 1.0 um).

     2.  Setting used during acquisition; especially, threshold trigger
(FSC or FL) and values thereof

     3.  Gating strategies

     4.  Sample collection and staining

Please respond to laurie_stojanovic@baxter.com with any information you
have that could help me.  I may also be reached at (847)270-4422.

Thanks!

Laurie Stojanovic



This archive was generated by hypermail 2b29 : Wed Apr 03 2002 - 11:53:22 EST