Re: simultaneous surface/DNA analysis

FACS_COPY@wehi.edu.au
Mon, 01 May 1995 13:41:56 +1000

On the subject of surface vs DNA analysis: unfortunately I have no
good suggestions to add but would like to comment on a couple of
suggestions made so far. The problem of PI (DNA) vs PE (surface) is
that the intense PI fluorescence leaks into the PE detector. Nothing
you do to the PI detector optical filters will change that; going to
a longer wavelength PI filter doesn't change the PI emission spectrum.
The PE detector will be subjected to just the same amount of
leakage. On the other hand, going to a *shorter* wavelength *PE
detector* optical filter may indeed take you into an area of better
PE-to-PI detection ratio.

| | < Frank Battye: battye@wehi.edu.au
\__/ <<<<< The Walter & Eliza Hall Institute
---------------!!<<<<<<<< Post Office, Royal Melbourne Hospital
/!!\ <<<<< Victoria 3050, Australia
o !! \ < ph: 61_3_345 2541, fax: 61_3_347 0852


Home Page Table of Contents Sponsors Web Sites
CD ROM Vol 2 was produced by staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone:(317) 494-0757; FAX (317) 494-0517; Web http://www.cyto.purdue.edu EMAIL robinson@flowcyt.cyto.purdue.edu