Re: Fluorescence intensity from large cells

Larry Seamer (larry@athena.unm.edu.unm.edu)
Thu, 6 Apr 95 09:30:30 GMT-0800

Dave Coder's response neglected one key aspect of Bjarne's original
question. It appears from the question that Bjarne is acquiring
*peak* fluorescence signals. However, Absolute fluorescence (e.g.
receptor number) is best measured on *integrated* fluorescence
signals, where the size of the cell (compared to the laser beam) is
not a factor. Newer FACScans and, I suspect, the FACSort have the
capacity to measure integrated fluorescence. With that caveat, I
agree with Dave's suggestions.


Home Page Table of Contents Sponsors Web Sites
CD ROM Vol 2 was produced by staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone:(317) 494-0757; FAX (317) 494-0517; Web http://www.cyto.purdue.edu EMAIL robinson@flowcyt.cyto.purdue.edu