From: Tony Ward (Tony_Ward@bdis.com)
Date: Fri Feb 11 1994 - 13:58:22 EST
REGARDING 3 color phenotyping with PerCP: - From in-house (BDIS) experiments it is possible to use PerCP on a FACStar Plus if laser power is reduced to 40 mW or less when using the 2 inch laser focusing lens. Sensitivity is not great but should be useable with bright markers. It may not be possible to run a large laser in a stable mode at low enough power to use PerCP. Focusing the laser to a large spot in the flow cell (eg using the 5 inch lens rather than 2 inch or using the horizontal expander with the 2 inch lens) will also reduce the laser intensity. Also using the low power compromises both FITC and PE sensitivity. Therefore, we recommend that you try lowering the laser power while observing the fluorescence histogram or dot plot for PerCP. Best Regards Tony Ward - BDIS