Re: CFSE Compensation Problem

From: Mark Kukuruga (kukuru@med.umich.edu)
Date: Thu Nov 15 2001 - 23:36:28 EST


Andrea,
You're running these samples on a Coulter XL . . . and that "problem" you're seeing
is unique (as far as I know) to the XL. (I have 2 Elite cell sorters, running the old
DOS Elite acquisition/analysis software -- they don't have this problem . . .)
When compensating, many of the negatives will bury against the baseline . . . a situation
even Coulter recognizes as problematic, as they now include an option in their (ACS's)
EXPO32 software to positively "offset" the negatives, placing them in the first decade
. . . and in view.
So, EXPO 32 is the only way to eliminate this "artifact."
MAK.


>>> Andrea Dewar <andrea.dewar@imvs.sa.gov.au> 11/15/01 19:28 PM >>>

Hi there.

I'm having a big problem with compensating cells double
stained with CFSE (FITC) and PE on our Coulter flow
cytometer. On a dot-plot, to bring single stained CFSE
cells into the 1st decade of the scale, 75% or more of
my cells slam against the axis before the remaining 25%
of cells are pulled into the first decade region.

When my cells are counter stained with a PE antibody,
the majority of negative cells are still slammed onto
the axis, so that if you don't look closely, you can't
see these cells and all events appear to be double
positive (whereas at least 25% of the cells are
actually negative).

This is a problem specific to the Coulter machine as
the same cells were run on a BD machine. The cells did
not slam into the axis when compensated on the BD
machine.

The level of compensation used on the Coulter machine
is not high (28.1) and the problem is not becuase the
cells are stained too brightly (the problem still
occurs when the cells are sitting at 2000).

Any uggestions would be thoroughly appreciated.

Andrea Dewar

PhD Student
Institute of Medical and Veterinary Science
Adelaide
South Australia
AUSTRALIA



This archive was generated by hypermail 2b29 : Sun Jan 05 2003 - 19:01:39 EST