Hi Claude, I think it's going to be tough to see the Cy7 from the PharRed using a -08 PMT. I'd recommend at least a -12, I'd prefer a -15. Are you seeing the APC from the PharRed- you might see a bit of that , which would verify that the Ab is working. Also, you may squeak out a tiny bit more light by using a short pass instead of going through a long pass to get to the Cy7. Good luck Peter Lopez The Aaron Diamond AIDS Research Center 212.448.5188 (office) 212.448.5159 (fax) 212.448.5190 or 5110 (lab) Claude Cantin <cantinc@uhnres.ut To: mug@Cytomation.com, oronto.ca> cytometry@flowcyt.cyto.purdue.edu Sent by: cc: owner-mug-cytomate Subject: no PharRed @frii.net 05/23/2001 06:06 PM Please respond to cantinc Hi All I have been unable to detect a PharRed signal from samples I have been given, and don't know if the problem is instrument or labeling error. MoFlo set up is as follows: Hammatsu 957-08 PMT all around Omega Optical Filter Set XF501 (consists of 710 DRLP, 670DF40, and 740EFLP) Melles Griot 35mW HeNe Red laser alignment seems fine: "Molecular Probes AlignPlus beads for 633" are almost off scale if no ND filter, at 399V, linear gain 1. BD's APC like Calibrites show excellent separation + / - ( >2 logs) 580V. Whether I measure APC in FL5 (remove 710LP dichroic, use 670/40BP) or FL6 (710LP plus 670BP), I obtain good APC signal. But no PharRed with the 710LP dichroic / 740EFLP in FL5, and no PharRed signal even if I remove the 710LP dichroic. BD Pharmingen Technical data sheet recommends a Hammatsu R3896 PMT for improved red sensitivity. Is the Hammatsu 957-08 deficient in the >740 range? If not, what's up? Thanks, Claude (See attached file: cantinc.vcf)
This archive was generated by hypermail 2b29 : Sun Jan 05 2003 - 19:01:20 EST