Hi Glenn, You ought to get the same results - one point is that since the sample size is high you need not count. You can just compare the -1 drop and +1 drop puddles until they contain the same number of outliers: speed. Counter mode has the extra 0.5 drop on each side, so it requires more settings to determine the delay. If you use Normal-R, there is no extra time added to the envelope, so 1 drop packets are reliable. I never said you would get different results, only that using Normal-R and the 3 puddle method is faster because the sample size is huge and the timing is 1.0 drop at 1 drop. I said "may be asking for trouble" using a 1.3 drop deflection envelope after a traditional Counter method. We usually time the sorter in couple of minutes, the traditional drop delay profile takes a lot longer. My other point, accurate means a sample size of 1500 +/- 100 is easier to spot with 1 drop envelopes than 50 +/- 1 with 1.5 drop envelopes, doesn't mean you'll get a wrong result - only that you'll need more deflection samples with different time increments to determine the delay. Joe Glenn Paradis wrote: > Hi Joe, > > I use counter mode to set drop delay all the time. I sort 50 beads at the > different delays and see which ones give me the best recovery. I heard > that you said that drop delays in counter vs normal-r would give different > results so I tested them and got exactly the same drop profile in both > modes. I did this on a TSO Vantage and my FACStar Plus. > > Just wondering about this normal-r vs counter mode difference?????? > > Glenn > > ************************************************* > Glenn Paradis > MIT Flow Cytometry Core Facility > E18-463 > Massachusetts Institute of Technology > Cambridge, MA 02139 > (617)253-6454 lab > (617)452-4132 office > gap@mit.edu > ************************************************* > > >Doing a drop delay profile on a Vantage using Counter mode and then sorting > >in Normal-R mode with a 1.3 drop deflection envelope may be asking for > >trouble. Counter mode adds 0.5 drops on either side of the drop envelope, > >thereby masking the prescision one needs to be able to clock down to a 1.3 > >drop window. I would suggest using the "3 puddle" method where you sort in > >Normal-R using 1.0 drop envelopes at: -1 drop, 0 drop, +1 drop from the best > >estimate of the delay running particles at ~ 500/second. Sort 1000 - 1500 > >events into each puddle and adjust the delay such that you get the same > >number of outliers in the -1 and +1 offsets (and the vast majority in the > >middle puddle). By using a real 1.0 drop envelope and using the larger sample > >size it is far easier to accurately time the instrument to within 0.1 drops. > >We have never used traditional drop delay profiles to time the sorters since > >the "3 puddle" method is faster and more accurate. > > > >The Vantage SE is a essentially a digital sort system, whereas the Vantage > >uses faster peak sense and hold circuits inherited from the FACStar Plus. > >Upgrading a Vantage to a Vantage SE makes a lot of sense if you want better > >sorting at "TurboSort" performance. > > > > > > Regards, > > > > Joe > >
This archive was generated by hypermail 2b29 : Wed Apr 03 2002 - 11:56:04 EST