          You should neither see a change in scatter, nor in 
          fluorescence.  Just avoid harsh changes in environment e.g. 
          pre-cool to ~20oC to slow down metabolism and try to sort 
          out of the medium into medium and use just PBS with no 
          additives as sheath. Isoton and other commercial sheath 
          solutions can kill the bugs fast (counterstain with PI at 1 
          to 10ug/ml).
          
          Make sure that all your media are not only sterile but also 
          filtered to avoid being drowned in noise, in particular when 
          rerunning your sorted sample.  If possible sort directly 
          onto agar plates as single cells if you have a cloning 
          attachment.
          