FACS 440 assistance: update.

Raymond B. Hester (rhester@jaguar1.usouthal.edu)
Fri, 12 Sep 1997 15:16:20 -0500 (CDT)

Dear group,

Our FACS 440 will not display dots or accumulating histograms on the data
display screen although pulses appear on the console pulse display (the
alpha numerics do appear, however, on the console data display).

We started swapping cards and boards from a used 440 we bought for this
purpose and replacement of the DMG board in the ND624 restored the dots -
for a time. The problem surfaced again the following day!

One other piece of information - our software runs on a VAX and on the
VAX monitor one can see the dots. We could perhaps operate in this
fashion except for the fact that it makes alignment almost impossible,
i.e.,
it's difficult to see the dots representing polystyrene beads on this
screen, and they don't seem to respond as quickly (no matter how you set
it up timewise) to adjustments in the stream or fluorescence pickup lens
as did the dots on the FACS console data display.

Does this give anyone a clue as to what has gone wrong?

Thanks again for any help.

Ray

++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++++